Access & Closure

How is Closure Managed with Impella® Heart Pumps?

Manual compression is the standard closure technique with Impella heart pumps.1 Key steps of the manual compression technique are:2

  • Sheath removal: Operator applies gentle pressure 1-2 cm above the access site and removes the sheath.
  • Hemostasis: Immediately, operator applies firm downward pressure for 15-30 minutes (5 minutes of full pressure, 5 minutes of 75% pressure, 5 minutes of 50% pressure and 5 minutes of 25% pressure) or until hemostasis is achieved.

 

 

View more resources

References

  1. Burzotta, F., et al. (2015). Int J Cardiol, 201, 684-691.
  2. Impella Certification Workbook 2018.

IMP-395

This information is intended for use by customers, patients, and healthcare professionals in United States only. We recognize that the Internet is a global communications medium; however, laws, regulatory requirements, and product information for medical products can vary from country to country. The product information included here may not be appropriate for use outside United States, and the information from other sites you visit may not be appropriate for use in United States.